Localisation

duration of localisation under local anaesthesia
metallic guidewire or magnetic seed depending on context
millimetre precision through imaging guidance
When sight and touch are not enough

The vast majority of breast lesions diagnosed today — early cancers, atypical hyperplasia, papillomas, microcalcifications — are non-palpable. They cannot be felt on clinical examination, are invisible on the skin surface, and cannot be seen with the naked eye once the skin is incised. They exist only on imaging: a 7 mm ultrasound nodule, a focus of microcalcifications on a mammogram, a suspicious magnetic resonance imaging (MRI) enhancement.
When these lesions must be surgically removed, a concrete problem arises: how does the surgeon know exactly where to operate? Without localisation, the surgeon would be operating blind — with a significant risk of missing the target or removing an excessive volume of healthy tissue. Pre-operative localisation is the technical solution to this problem: it consists in placing, within the breast, a physical marker that the surgeon can identify in theatre to guide the procedure.
Key point: pre-operative localisation turns an invisible and impalpable lesion into a tangible target for the surgeon. It is essential for any surgery of a non-palpable lesion, whatever its nature — cancer, at-risk lesion (atypical ductal hyperplasia (ADH), lobular carcinoma in situ (LCIS), radial scar), papilloma, or a simple benign nodule to be removed.
When is it indicated?
Localisation is indicated as soon as a breast lesion to be operated on is not clinically accessible — i.e. cannot be found under the surgeon's fingers in theatre. In practice, this concerns: small breast cancers detected at screening, atypical ductal hyperplasia (ADH) and other B3 lesions, papillomas, suspicious foci of microcalcifications (Breast Imaging-Reporting and Data System (BI-RADS) 4 or 5), small ultrasound nodules that are deep or poorly defined, and certain lesions detected only on MRI with no ultrasound correlate — which then require MRI-guided localisation, a rarer and more complex procedure.
Source: SENORIF 2025-2026 · French Society of Breast Oncology & Pathology (SFSPM)
The biopsy clip: the target left behind

Before any surgery, there has almost always been a biopsy. During this biopsy — ultrasound-guided core biopsy for a nodule, stereotactic vacuum-assisted excision for microcalcifications, MRI-guided biopsy for an MRI-only lesion — the radiologist samples one or several fragments of breast tissue. At the end of the procedure, a small metal clip of a few millimetres, usually titanium, is systematically deposited at the exact site of sampling.
The most widely used clips are Twirl, Tumark, and UltraClip. They are biologically inert, painless, and fully compatible with mammography, ultrasound and breast MRI. They can remain in place indefinitely without any drawback.
Why is this clip essential?
After a biopsy — especially for microcalcifications or a small lesion — it is very common for the target to have entirely disappeared from imaging: the biopsy itself removed it all. If no clip had been placed, it would then be impossible to find precisely where the sampling had been done.
The clip therefore acts as a lasting anatomical reference point. It serves two purposes:
If no surgery is decided (benign result, surveillance), the clip allows accurate radiological follow-up. At each control mammogram, the radiologist can review the exact biopsied zone and confirm its stability.
If surgery is decided, the clip serves as the target for pre-operative localisation. The localisation marker (metallic guidewire or magnetic seed) is placed at the metal clip — pointing to it. By removing the marker and the clip, the surgeon mechanically removes the area of interest.
A metal clip is always placed during a biopsy of a non-palpable lesion — it is good practice that protects the patient by keeping a physical trace of the biopsied zone, whatever the histological result and whatever the subsequent decision.
Source: SENORIF 2025-2026 · French National Authority for Health (HAS)
Metallic guidewire or magnetic seed

Two main pre-operative localisation techniques are currently available. Both are validated and effective; they differ mainly in logistics and patient comfort. The choice depends on availability at the radiology centre, lesion topography, care pathway organisation and patient preference.
Metallic guidewire
A thin metal wire with one end anchored in the breast at the biopsy clip and the other end exiting the skin. Placed the evening before or the morning of surgery, under ultrasound or stereotactic guidance.
A few hours (same day) to 24 h maximum
Magnetic seed
A small magnetic marker of about 5 mm (Magseed, Localizer, Sirius, Savi Scout) placed at the metal clip. Detected intra-operatively by a dedicated magnetic probe.
Several days to several weeks (even months)
One goal, two ways to get there
The goal is strictly the same in both cases: enable the surgeon to precisely locate the lesion in theatre. The oncological outcome is equivalent — it is not a question of surgical quality, but of organisation. The question is therefore not "which technique is best?" but "which is most suited to this patient, at this centre, for this lesion?".
Source: SFSPM · SENORIF 2025-2026
You have just been told that a small, non-palpable breast lesion needs surgery and you would like to be operated on by a breast cancer surgeon.
Dr Zeitoun sees you for a second opinion on the localisation technique — metallic guidewire or magnetic seed — and on the surgical strategy.
Metallic guidewire

The metallic guidewire, also called wire localisation or guidewire, is the historical technique for pre-operative breast localisation. In use since the 1970s, it remains today the most widely available method — available at every breast radiology centre, reliable, well-established, inexpensive.
How it works
A very thin metal wire (not unlike a fishing line in appearance) is introduced into the breast through a carrier needle, under ultrasound or mammographic guidance. The inner end of the wire carries a small hook which anchors in breast tissue at the biopsy clip. The needle is then withdrawn and the wire stays in place, with its outer end exiting the skin — covered by a sterile dressing.
Practical workflow
The day before or the morning of surgery
The patient arrives at the radiology practice. If localisation is done the day before, no fasting required.
If done the same morning, she arrives fasting, typically between 7:30 and 9:00 am. The specialist breast radiologist reviews the images, discusses the procedure and prepares the patient.
Placement under local anaesthesia
Antiseptic cleansing, local anaesthesia in the skin and breast tissue, placement of the metallic guidewire under ultrasound or stereotactic guidance. Duration: 15 to 20 minutes.
Transfer to theatre
X-ray check, dressing, and direct transfer to theatre with the images. The patient remains fasting and does not remove the dressing before surgery.
Advantages and disadvantages
Universal availability: all breast radiology practices offer this technique.
Proven reliability: 50 years of experience, very high success rate.
Cost: inexpensive, well reimbursed by French national health insurance. Note: in the private sector, additional fees may apply depending on the radiology practice.
Constraining logistics: if localisation is done on the same morning, the patient must arrive fasting at the radiologist, then at theatre, with tight timing. If done the day before, no fasting required for placement — but the patient must attend twice (radiologist the day before, theatre the next day).
Wire exiting the skin: uncomfortable, anxiety-inducing for some patients, theoretical risk of displacement.
Time dependency: surgery must follow quickly. If surgery is postponed, the metallic guidewire must be removed and replaced later.
Source: SFSPM · HAS
The magnetic seed: Magseed, Localizer, Sirius

Developed in the 2010s, magnetic seeds represent a significant evolution in pre-operative localisation. Marketed under various names — Magseed (Endomag), Localizer (Hologic), Sirius, Savi Scout (Cianna Medical) — they share a common principle: a small magnetic metal capsule detected intra-operatively by a dedicated probe. These devices are being progressively deployed, with variable availability across radiology centres.
How it works
The magnetic seed is a metal capsule of about 5 mm in length, made of medical-grade stainless steel. It is introduced into the breast through a carrier needle, under ultrasound or mammographic guidance, and deposited immediately at the biopsy metal clip.
The needle is then withdrawn — unlike the metallic guidewire, nothing exits the skin. The seed rests silently in place until surgery.
In theatre, the surgeon uses a magnetic detection probe (e.g. Sentimag for Magseed) that emits an audible signal proportional in intensity to proximity to the seed. The closer the probe, the stronger the signal — the surgeon follows the signal to locate the target precisely, then removes the tissue surrounding the magnetic seed and the biopsy metal clip.
Practical workflow
Several weeks before
Standard appointment at the radiology practice, no fasting, at a convenient time for the patient. Placement of the magnetic seed under local anaesthesia, with ultrasound or mammographic guidance.
Normal life in between
The patient resumes normal activities immediately after placement. The seed is entirely painless, invisible, and does not interfere with movement, sleep, sport or daily life.
Surgery day at theatre
Direct arrival at theatre, fasting, with no intermediate step at the radiologist. The surgeon uses the magnetic probe to locate the seed during the operation.
Advantages and disadvantages
Decoupled logistics: placement several weeks before surgery, appointment independent of theatre.
Patient comfort: no wire exiting the skin, no anxiety on surgery day linked to localisation, no prolonged fasting.
Flexibility: surgery can be postponed without removing the seed, which remains active for several weeks.
Surgical precision: the probe allows fine guidance of the surgical gesture, often judged more comfortable by experienced operators.
Diagnostic safety: placement several weeks before surgery allows time for a full radiological review and, if needed, for additional biopsies in the same or the contralateral breast before the definitive surgical procedure.
Limited availability: not all radiology centres are equipped — deployment ongoing.
Cost: today fully reimbursed by French national health insurance and well covered. Note: in the private sector, additional fees may apply depending on the radiology practice.
MRI artefact: the seed generates a localised artefact on follow-up MRI (blur zone around the seed) — usually not a problem in practice.
Source: Endomag (Magseed), Hologic (Localizer), clinical studies 2018-2024
The specimen radiograph

A step often unknown to patients, but systematic in any localisation-guided surgery: the specimen radiograph. It is an essential verification step, performed on the spot, before skin closure.
What is it?

Once the surgeon has removed the fragment of breast tissue theoretically containing the lesion, the localisation marker and the metal clip, the specimen is sent to the radiology room adjacent to theatre (or X-rayed directly in theatre, depending on the centre), before being transmitted to the pathology laboratory. A simple mammogram of the fragment is performed. This specimen radiograph provides several decisive pieces of information within minutes.
Presence of the metal clip
The biopsy metal clip is visible on the specimen X-ray. If so, the target has been reached: the surgeon has removed the area of interest.
Presence of the localisation marker
The metallic guidewire or magnetic seed must also be present in the specimen — additional confirmation that the targeted zone has been retrieved.
Presence of microcalcifications
When the initial target was a focus of microcalcifications, their presence on the specimen X-ray confirms that the entire focus has been removed.
Position within the specimen
The position of the clip and microcalcifications relative to the margins of the specimen gives a first appreciation of margin status — completed by the histopathological analysis.
And if the clip is not on the X-ray?
This is a rare but anticipated situation. If the specimen X-ray shows neither the clip nor the target microcalcifications, the target has not been reached — the surgeon has not removed the correct zone.
In this case, a further resection is performed during the same operative time: the surgeon re-aims with the localisation marker, removes more tissue, and an X-ray of the new specimen is repeated until verification. This control step is an integral part of procedural safety — it prevents late discovery that the target has not been reached.
In practice: the specimen radiograph is systematic after any localisation-guided excision. It guarantees to the patient and the team that the target has indeed been reached — it is a quality standard in modern breast surgery.
Source: SFSPM · SENORIF 2025-2026
Your localisation and your breast surgery are scheduled soon.
A pre-operative consultation with Dr Zeitoun lets you concretely anticipate the procedure, ask all your questions and leave with a clear plan.
Your questions, answered
The answers below are drawn from daily clinical practice and current guidelines.
Is localisation painful?
No, or very little. Localisation is performed under local anaesthesia by the specialist breast radiologist. A small injection is given into the skin and breast tissue before placement of the metallic guidewire or magnetic seed. The procedure takes 10 to 20 minutes depending on technique and location. You may feel pressure or a slight pulling sensation with the metallic guidewire, but pain is very limited. A mild haematoma is possible, resolving spontaneously in a few days.
Which technique is best: metallic guidewire or magnetic seed?
Neither is superior to the other in terms of oncological outcome. Both techniques allow the surgeon to locate the lesion precisely — which is exactly the goal. The difference lies in logistics and comfort: the magnetic seed significantly simplifies the pathway by allowing placement several weeks before surgery, with no guidewire exiting the skin. The metallic guidewire remains very reliable and universally available. The final choice depends on availability at the radiology centre, lesion topography, and your preferences.
How long can a magnetic seed stay in place?
Depending on the device, from several days to several months. Available devices (Magseed, Localizer, Sirius, Savi Scout) allow placement sufficiently far in advance of surgery to fully decouple localisation logistics from theatre. This flexibility is one of the major strengths of the magnetic seed: it allows surgery scheduling without any localisation-related constraint, and also permits a full radiological review or additional biopsies in the same or the other breast before the definitive surgical procedure.
Does the magnetic seed trigger security gates?
No. The magnetic seed is very small (about 5 mm) and its magnetic signature is too weak to trigger airport gates, shop anti-theft devices or everyday detectors. You can travel and pass security checks without any particular constraint.
What if my breast is small or the lesion very deep?
Both techniques adapt to all breast morphologies and lesion depths, with a few nuances. The metallic guidewire can be trickier to position in a very dense breast or for a very deep lesion — but the specialist breast radiologist adapts the technique case by case. The magnetic seed is more versatile in this context. When difficulty is anticipated, discussion between surgeon and radiologist before localisation allows the best strategy to be planned.
Can the metallic guidewire move before surgery?
The theoretical risk exists — that is why the metallic guidewire is placed shortly before surgery (a few hours to 24 hours maximum). In practice, the modern metallic guidewire has an anchoring hook that holds it firmly in place. It is protected by a thick dressing, and it is recommended to limit sudden movements and avoid carrying heavy loads between placement and surgery. Significant displacement is rare; if the position seems uncertain, a new X-ray is taken before the procedure.
Can there be several localisations in the same breast?
Yes. When several non-palpable lesions must be removed during the same procedure, multiple localisations can be placed — one per lesion. Metallic guidewires can be multiple, as can magnetic seeds, although some probes have resolution limits between two very close seeds. The radiologist and surgeon discuss the strategy in advance to optimise positions.
Is localisation reimbursed?
Yes, both techniques are reimbursed by French national health insurance. The metallic guidewire is well reimbursed — a historical and inexpensive technique. The magnetic seed (Magseed, Localizer, Sirius) is today fully reimbursed by French national health insurance and well covered. Note however: in the private sector, additional fees may apply — whether for metallic guidewire or magnetic seed — depending on the radiology practice and its convention with national health insurance. Always ask the practice for a prior quote stating the tariff and any out-of-pocket cost.
What happens to the biopsy metal clip after surgery?
It is removed with the surgical specimen. This is in fact one of the explicit goals of surgery: to remove en-bloc the localisation marker (metallic guidewire or magnetic seed), the biopsy metal clip and the surrounding lesional tissue. The specimen radiograph confirms its presence in the resected fragment. After surgery, there is therefore no clip left in the breast.
Do I have to remove the metallic guidewire if my surgery is postponed?
Yes, if the delay exceeds a few hours. The metallic guidewire is designed to stay in place for a very short time. If for any reason surgery is cancelled or postponed beyond 24-48 hours, the metallic guidewire must be removed by the radiologist, then repositioned before the new surgical date. That is one of the reasons why, when possible, the magnetic seed offers more flexibility: it can stay in place for several weeks without drawback, even if surgery is postponed.
What are Dr Zeitoun’s fees?
Dr Zeitoun practises in Sector 2 (non-OPTAM) and charges fees above the French Social Security rate for all consultations and procedures. Most international patients are not covered by French Social Security. A detailed quote is given during the consultation, before any procedure. If you are covered, the Assurance Maladie reimburses on the basis of the Social Security rate, and your health insurance may cover all or part of the additional fees, depending on your policy.
A written quote is given beforehand. Sector 2, with fees above the standard health-insurance rate.
Surgery for a non-palpable breast lesion has been proposed to you and you would like it to be performed by a breast cancer surgeon.
Dr Zeitoun coordinates your localisation with the specialist breast radiologist, and chooses with you the most suitable technique — metallic guidewire or magnetic seed.
Sources
Institutional sources and scientific literature
This page draws on French and international breast oncology guidelines, the recommendations of learned societies, and the medical literature indexed on PubMed.
- SENORIF 2025-2026 — French national breast oncology reference. Management of non-palpable breast lesions and pre-operative localisation. 2025-2026 edition.
- SFSPM — French Society of Breast Oncology & Pathology. Recommendations for the management of non-palpable lesions. sfsenologie.com
- HAS — French National Authority for Health. ACR BI-RADS classification, good practice in breast biopsy and guided surgery. has-sante.fr
- Harvey JR, Lim Y, Murphy J, et al. Safety and feasibility of breast lesion localization using magnetic seeds (Magseed): a multi-centre, open-label cohort study. Breast Cancer Res Treat. 2018;169(3):531-536. PMID: 29445940
- Hayes MK. Update on Preoperative Breast Localization. Radiol Clin North Am. 2017;55(3):591-603. PMID: 28411682
- Hersi AF, Pistiolis L, Dustler M, et al. Optimizing Preoperative Breast Localization: A Comparison of Magseed and Wire-Guided Localization. Ann Surg Oncol. 2023;30(6):3353-3361.
- Rubio IT, et al. European guidelines for the management of breast lesions of uncertain malignant potential (B3 lesions). Eur J Surg Oncol. 2024.
- Endomag. Magseed — Technical notice and conditions of use. endomag.com
You would like to be operated on for a breast lesion localised before surgery, or to get a second opinion on your file.
Dr Zeitoun sees patients at his Paris 8th practice and at Clinique Hartmann (Neuilly-sur-Seine); a second opinion is possible on your records.